The spectrum appears in the Empower 3 Review window but not in the Spectrum Index Report; shows error "...contains information that doesn't match the data being reported" - WKB195829
Article number: 195829
SYMPTOMS
- The spectrum appears in the Review window but not in the Spectrum Index Report
- Error: "...contains information that doesn't match the data being reported"
ENVIRONMENT
- Empower 3
CAUSE
The processing method type is "LC".
FIX or WORKAROUND
- After the New Processing Method wizard opens, select "PDA" for processing type on the second page and create a new processing method.
- Register the new processing method in the method set and perform processing.
- Preview the created result.
ADDITIONAL INFORMATION
id195829, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP