Skip to main content
Waters

Empower error: ORA-01426, Numeric Overflow - WKB18471

Article number: 18471

SYMPTOMS

  • No result is produced, regardless of how the sample is processed
  • Error observed in Empower message center: ORA-01426, Numeric Overflow

ENVIRONMENT

  • Empower 3 FR3
  • Processing method has a peak width of 1 second or less

CAUSE

Unknown; a defect (PCS# 38126) has been opened.

FIX or WORKAROUND

Set a peak width in the processing method that is greater than 1.

ADDITIONAL INFORMATION

Fixed in Empower 3 FR4

id18471, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP

 

Not able to find a solution? Click here to request help.