SYMPTOMS

  • Error: ORA-12528: TNS Listener: All appropriate instances are blocking new connections observed when logging in to Empower 3
  • A reboot does not resolve the error
  • Restarting the Oracle services does not resolve the error

ENVIRONMENT

  • Empower 3 Personal

CAUSE

The adump folder is too large.

FIX or WORKAROUND

  1. Stop the Oracle services.
  2. Rename the adump folder.
  3. Create a new adump folder.
  4. Copy the adx.txt file from the old folder into the new folder.
  5. Restart the Oracle services.

ADDITIONAL INFORMATION

C:\Empower\Oracle\admin\Empower\adump