Waters Database Manager (WDM) in Empower or UNIFI / Waters Connect displays the error "ORA-27369: job of type EXECUTABLE failed with exit code 15 . . ." when configuring Backup Policy - WKB16381

Article number: 16381

SYMPTOMS

  • Configuring Backup Policy in Waters Database Manager (WDM) in Empower generates the error "ORA-27369: job of type EXECUTABLE failed with exit code:"
  • image.png
  • Configuring Backup Policy in Waters Database Manager (WDM) in Empower generates the error ORA-27369: job of type EXECUTABLE failed with exit code: The system cannot find the drive specified.

ENVIRONMENT

  • Empower 3 FR4
  • UNIFI Workstation
  • UNIFI Network
  • waters_connect

CAUSE

The oraclejobs OS user is not in the Backup Operators group.

The local group "Backup Operators" is not present in the "Log on as batch job" local security policy setting.

FIX or WORKAROUND

  1. Add the oraclejobs user to the Backup Operators group.
  2. Modify the Local Security Policy "Log on as batch job" to include the local group "Backup Operators" or explicitly list the oraclejobs user.

ADDITIONAL INFORMATION

clipboard_ef8ab19f088c8916899a6dc1dd40eeb2d.png

Note: In some cases you may need to logout of the Windows system or reboot for the above actions to take effect.