ORA-27369 error: Job of type executable failed with exit code: Incorrect function - WKB16254
Article number: 16254
SYMPTOMS
- Trying to run WDM FR4 backup but unable to submit job
ENVIRONMENT
- Empower 3 FR4 Enterprise Server
CAUSE
Empower 3 Enterprise Server cannot be installed on a Domain Controller.
There are seven BAT files that should be copied to the C:\Empower\tmp\scripts folder.
Copying did not occur.
FIX or WORKAROUND
- Install Empower 3 Enterprise Server on a non-Domain Controller.
- Copy the seven script files from C:\Empower\Oracle\WDM_InstallSupport to C:\Empower\tmp\scripts.
- The Empower\tmp\scripts are located on the Oracle application drive.
- The seven script files are confrman.bat, PostBackupCmd.bat, Rundpexport.bat, Rundpimport.bat, runrman.bat, runsqlscript.bat, and show_config.bat.