Peaks are integrated but not labeled in the Empower Review window - WKB15949
Article number: 15949
SYMPTOMS
- Peaks are integrated but not labeled when reviewing the data
- Retention times are as listed in the components table
- No errors are observed in the message center
ENVIRONMENT
- Empower 3 Personal
- H-Class with QDa
- Internal standard
CAUSE
The wrong channel is designated in the processing method.
FIX or WORKAROUND
Select the correct channel.
ADDITIONAL INFORMATION