"COM Fatal error" on Agilent 7890 GC using Empower - WKB15676

Article number: 15676

SYMPTOMS

  • When clicking on Run Samples and selecting system, the error "COM Fatal error" appears
  • Node properties show instrument status "No" for GC and Headspace
  • No changes to DCOM or any Windows settings
  • Issue involves only one LAC/E

ENVIRONMENT

  • Empower 3 Feature Release 1 (FR1) - client server
  • Agilent 7890A
  • Agilent 7697A Headspace

CAUSE

Loss of instrument connection

FIX or WORKAROUND

  1. Remove the 7890A and the 7697A from the DHCP server window.
  2. Shut down the LAC/E and the system.
  3. Power-up the LAC/E.
  4. Power-up the system.

ADDITIONAL INFORMATION

See article WKB62833 on Split injection technique

See article WKB62831 on Splitless injection technique

See article WKB62933 on PTV injection technique

id15676, 7890 GC, 7890 GCDET, 7890BGC, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, HEADSPACE, SUP