Missing Vial or Wrong Plate Error - WKB1527

Article number: 1527

SYMPTOMS

  • Missing Vial or Wrong Plate Error
  • Sample set defaults to V:1 position

ENVIRONMENT

  • ACQUITY Classic

CAUSE

  • The plates are not defined prior to creating a sample set

FIX or WORKAROUND

  1. Define plates: From the Run Samples Window Select Customize > Defaults > Plate Define > Plate Type Name > ANSI-48Vial2mLHolder for Plate 1 & 2 > OK > OK.
  2. Save preferences: File > Save Preferences.
  3. Create a new sample set.

ADDITIONAL INFORMATION

 

id1527, UPBINARY