2424 communication fails when installed with ACQUITY Arc - WKB14754
Article number: 14754
SYMPTOMS
- When system with 2424 is accessed in Run Samples, Control Panel freezes for various components (see below)

ENVIRONMENT
- Empower 3 Enterprise, English
- Installed on LAC/E
- DP 2017 R1
- Installed with ACQUITY QSM-R, ACQUITY FTN-R
CAUSE
Installation folder directory is incorrect. (SCR# 35552)
FIX or WORKAROUND
- Change Region and Language to English-US.
- Change directory path from C:\Empower\Instruments\HTML\W2424\en to C:\Empower\Instruments\HTML\W2424\~en