With Empower, gradient start before injection in QSM-R instrument method is not operating correctly - WKB14666
Article number: 14666
SYMPTOMS
- Gradient start before injection in QSM-R instrument method is not operating correctly
- When set, regardless of changing volume or time, the chromatogram for gradient start before injection looks like the gradient start at injection
- Gradient start after injection looks OK
ENVIRONMENT
- Empower 3 Feature Release 3 (FR3)
- ACQUITY Arc
- Waters Driver Pack 2017 Release 1 (DP 2017 R1)
CAUSE
The issue with gradient start before the injection only occurs if you have a solvent select valve.
This has been verified by test and development.
FIX or WORKAROUND
The issue has been added to the defect database as PCS-35595, and it will be addressed in the next ICS for ACQUITY Arc.
This has been transfered to CRI-4609
ADDITIONAL INFORMATION
