"OCI_Invalid_Handle" error message in Empower after creating a new project - WKB13635
Article number: 13635
SYMPTOMS
- "OCI_Invalid_Handle" error message after creating a new project
- Cannot open the new project
- New project had 0 MB of free tablespace
ENVIRONMENT
- Empower 3
CAUSE
The project created had view filters, custom fields, methods, and preferences copied from another project. Those copies took up 50 MB of tablespace, resulting in 0 MB of free tablespace.
FIX or WORKAROUND
Increase the amount of tablespace from the default of 50 MB when creating the project.