Error: Database Connection ORA 12545: Connect failed because target host or object does not exist - WKB1327
Article number: 1327
SYMPTOMS
- Getting message "Database Connection ORA 12545: Connect failed because target host or object does not exist at login"
ENVIRONMENT
- Empower 3 Enterprise
CAUSE
Confirm the HOST entry in the Empower client TNSNAMES.ora file (located under C:\Empower\Oracle\Oracle11gClient_2\network\admin). Could be:
- Incorrect IP address listed
OR
- Incorrect server name listed (may require servername.domain.com OR just servername)
FIX or WORKAROUND
- Correct HOST entry in the Empower client's TNSNAMES.ora file