Empower client error: Could not connect to requested node - WKB13273
Article number: 13273
SYMPTOMS
- Empower 3 client showed this error when viewing node properties on the Empower server
ENVIRONMENT
- Empower 3
CAUSE
The Empower client PC was not logged in to the domain.
FIX or WORKAROUND
- Log out of Windows on the Empower client.
- Log in to the domain instead of using a local user account.
- Log in to Empower.