ACQUITY UPLC FLR Detector is not communicating in Empower - WKB126084
Article number: 126084
SYMPTOMS
- ACQUITY UPLC FLR Detector is not communicating in Empower
- ACQUITY UPLC FLR populates in DHCP
- Can ping IP Address of ACQUITY UPLC FLR Detector
- ACQUITY UPLC FLR Detector shows up as Yes in Node Properties
- ACQUITY UPLC FLR Detector control panel is blank in the Run Samples window
ENVIRONMENT
- ACQUITY UPLC FLR
- Empower 3
- Feature Release 5
- Driver Pack 2019 Rev 2
CAUSE
Antivirus is blocking communication.
FIX or WORKAROUND
- Temporarily disable antivirus and reboot the detector and computer.
- In the Run Samples window, verify that the detector is communicating properly.
- Turn antivirus back on and set the antivirus exclusions per WKB16341.