2998 control panel is empty in the Empower Run Samples window - WKB125152
Article number: 125152
SYMPTOMS
- 2998 control panel is empty in the Empower Run Samples window
- Monitoring the baseline is possible
- Injections are possible
- If AcquityConsole.exe is launched from File Explorer, the 2998 shows up correctly in that window
- No firewall or antivirus is running
- COM security is set up to give "Everyone" full control for both "Access Permissions" and "Launch and Activation Permissions"
- Restarting both the PC and the system does not help
ENVIRONMENT
- Empower 3 Personal
CAUSE
Unknown
FIX or WORKAROUND
- Close the Run Samples window.
- In Empower's Configuration Manager window, take the system offline, and then bring it back online.
- Open the Run Samples window to verify that the control panel is now populated.
ADDITIONAL INFORMATION