Agilent GC "Server Busy" error in ICF Support Layer v3.4 and Instrument Control Framework 3.0 - WKB124100

Article number: 124100

SYMPTOMS

  • Sample set does not start
  • "Server Busy" error appears with Switch to / Retry button

ENVIRONMENT

  • Empower 3
  • Agilent ICF Support Layer v3.4
  • Agilent Instrument Control Framework 3.0

CAUSE

Column parameters are missing.

FIX or WORKAROUND

  1. Specify all required column parameter information in the instrument method in Configuration / Columns.
    Select the column and click on Catalog to enter all information in "Additional information (Optional)".
    clipboard_ecbae76444fb2599c79eca49ff9fc0111.png
  2. Refer to the diagnostic messages in the console; if necessary, give setup instrument method twice.

ADDITIONAL INFORMATION

This issue was raised as CRI-2932.

It is fixed fixed in Agilent ICF 3.0 Update 2, part of ICF Support Layer v3.5.

Other server busy error with Agilent ICF Support Layer 3.4 : 
Instrument Method if Installed GC Column "Min Temperature" is in Negative (-) number it will lead to Server Busy error with Agilent ICF v3.4 in Agilent GC - WKB195431

 

 