Incorrect standard amount calculated with Empower - WKB12359
Article number: 12359
SYMPTOMS
- Incorrect standard amount calculated in Empower
ENVIRONMENT
- Empower 3
- Review window
- Amount tab
CAUSE
Incorrect calibration fit type. Use a single-level calibration such as linear through zero or response factor, not a multi-level calibration fit.
FIX or WORKAROUND
Change the calibration fit type to Linear through zero.