How to configure Empower to access the extra 10-vial plate for Agilent G4226A - WKB12195

Article number: 12195

OBJECTIVE or GOAL

  • Inject from the extra 10-vial plate on the Agilent G4226A autosampler.

ENVIRONMENT

  • Empower 3
  • Agilent Autosampler G4226A
  • Agilent Multisampler G7167B
  • Agilent ICF Support  version 2.2 and above

PROCEDURE

  1. In Configuration Manager, select Plate Types, right-click on the table, and select Import from Text.
  2. Browse to the vialbar.txt file on the Agilent ICF media (under \AgilentPlatesForImport\Empower 3\En-US).
  3. When using the plate, assign it as plate 0 (zero). When assigning usage of the plate, the syntax of the plate is 0:1 to 0:10.  

ADDITIONAL INFORMATION

See the attached PDF for details on configuration and usage (example from Agilent).

 

PLEASE NOTE THE FOLLOWING FOR MULTISAMPLER G7167B

This procedure also applies for set up of extra 5  vial plate  for  Multisampler G7167B but in this case the direction is Vertical Discontinuous

 