How to configure Empower to access the extra 10-vial plate for Agilent G4226A - WKB12195
Article number: 12195
OBJECTIVE or GOAL
- Inject from the extra 10-vial plate on the Agilent G4226A autosampler.
ENVIRONMENT
- Empower 3
- Agilent Autosampler G4226A
- Agilent Multisampler G7167B
- Agilent ICF Support version 2.2 and above
PROCEDURE
- In Configuration Manager, select Plate Types, right-click on the table, and select Import from Text.
- Browse to the vialbar.txt file on the Agilent ICF media (under \AgilentPlatesForImport\Empower 3\En-US).
- When using the plate, assign it as plate 0 (zero). When assigning usage of the plate, the syntax of the plate is 0:1 to 0:10.
ADDITIONAL INFORMATION
PLEASE NOTE THE FOLLOWING FOR MULTISAMPLER G7167B
This procedure also applies for set up of extra 5 vial plate for Multisampler G7167B but in this case the direction is Vertical Discontinuous