Skip to main content
Waters

How to configure Empower to access the extra 10-vial plate for Agilent G4226A - WKB12195

Article number: 12195

OBJECTIVE or GOAL

  • Inject from the extra 10-vial plate on the Agilent G4226A autosampler.

ENVIRONMENT

  • Empower 3
  • Agilent Autosampler G4226A
  • Agilent Multisampler G7167B
  • Agilent ICF Support  version 2.2 and above

PROCEDURE

  1. In Configuration Manager, select Plate Types, right-click on the table, and select Import from Text.
  2. Browse to the vialbar.txt file on the Agilent ICF media (under \AgilentPlatesForImport\Empower 3\En-US).
  3. When using the plate, assign it as plate 0 (zero). When assigning usage of the plate, the syntax of the plate is 0:1 to 0:10.  

ADDITIONAL INFORMATION

PLEASE NOTE THE FOLLOWING FOR MULTISAMPLER G7167B

This procedure also applies for set up of extra 5  vial plate  for  Multisampler G7167B but in this case the direction is Vertical Discontinuous

clipboard_e97c3fe659bddca4f898d69a09615c28b.png

id12195, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP

Not able to find a solution? Click here to request help.