Skip to main content
Waters

How to configure Empower to access the extra 10-vial plate for Agilent G4226A - WKB12195

Article number: 12195

OBJECTIVE or GOAL

  • Inject from the extra 10-vial plate on the Agilent G4226A autosampler.

ENVIRONMENT

  • Empower 3
  • Agilent autosampler G4226A
  • Agilent ICF Support v2.2

PROCEDURE

  1. In Configuration Manager, select Plate Types, right-click on the table, and select Import from Text.
  2. Browse to the vialbar.txt file on the Agilent ICF media (under \AgilentPlatesForImport\Empower 3\En-US).
  3. When using the plate, assign it as plate 0 (zero). When assigning usage of the plate, the syntax of the plate is 0:1 to 0:10.  

ADDITIONAL INFORMATION

See the attached PDF for details on configuration and usage (example from Agilent).

Not able to find a solution? Click here to request help.