What are the ICF driver recommendations for interoperability? - WKB121852

Article number: 121852

ENVIRONMENT

  • Empower 3 Enterprise
  • Agilent Instrument Control Framework
  • Agilent LC
  • Agilent GC
  • Agilent 6890
  • Agilent 7890
  • Agilent 8890
  • Agilent HS7697A
  • Agilent G18888

ANSWER

  • From ICF SL 3.2 with 2.6 Update 2 and above, when using Agilent gas chromatography (GC) with or without headspace (HS) modules, you must install the same ICF Support Layer software version on all Empower Citrix servers, clients, and LAC/E systems.
  • When using Agilent liquid chromatography (LC) modules, the Agilent ICF SL drivers installed (until otherwise noted) on Empower Citrix servers and clients are interoperable with LAC/E modules installed with ICF Support Layer 3.0 (A.02.04 Agilent ICF) and later in Empower.