Apex Tracking integration not integrating small peaks in Empower - WKB120889
Article number: 120889
SYMPTOMS
- Apex Tracking integration not integrating small peaks in Empower
- Integration window is set to include small peaks
ENVIRONMENT
- Empower 3
CAUSE
Apex Tracking automatically set the peak width too high that it excludes the small peaks from integrating
FIX or WORKAROUND
- Use the Event Type "Set Peak Width (sec)" in the Processing Method and enter a smaller value
ADDITIONAL INFORMATION
id120889, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP