SYMPTOMS

  • Apex Tracking integration not integrating small peaks in Empower
  • Integration window is set to include small peaks

ENVIRONMENT

  • Empower 3

CAUSE

Apex Tracking automatically set the peak width too high that it excludes the small peaks from integrating

FIX or WORKAROUND

  1. Use the Event Type "Set Peak Width (sec)" in the Processing Method and enter a smaller value 