In Empower 3, some channels are not processed - WKB120667

Article number: 120667

SYMPTOMS

  • Some channels are not processed, even if the sample set is run with "Run and Process"/"Run and Report" specified
  • Message center shows error "Oracle Fetch error with schema W_XXXXXXXXXX trying to get next valid ID, ID returning = 0"
  • Unprocessed channel data can be reviewed

ENVIRONMENT

  • Empower 3 Workgroup
  • Empower 3 Enterprise

CAUSE

Temporary buffering occurred in LAC/E during the sample set run.

FIX or WORKAROUND

  1. Confirm communication between server and LAC/E.
  2. Reprocess the sample set.