With Empower 3 and Agilent 7697A Headspace, are the vial filling modes "Default: constant flow rate up to specified pressure" and "User settings" > "Constant flow rate up to specified pressure" the same? - WKB118881
ENVIRONMENT
- Empower 3
- Agilent 7697A
- Headspace
- ICF 3.2
- ICF 3.3
ANSWER
Yes, they are the same.
If you select "Constant flow rate up to specified flow rate" in "User settings" and save the instrument method with the default "Filling flow rate", "Filling pressure", and "Retention time", "Default: Constant flow rate up to specified pressure" is automatically set. It returns to "Flow rate". When using "Filling flow rate", "Filling pressure", and "Retention time" in the initial setting, it is recommended that you use "Initial setting: Constant flow rate up to the specified pressure" instead of "User setting".
The following two test conditions are the same.
■ Condition A
Vial filling mode: "Default: constant flow rate up to specified pressure": filling pressure (15psi)
■ Condition B
Vial filling mode: "User setting" "Constant flow rate up to specified pressure": Filling flow rate (50 mL / min), Filling pressure (15 psi), Hold time (0.1 min)
■ Condition A

■ Condition B