Empower 3 Enterprise shows "System Idle - Instrument failure" in Run Samples window - WKB118379
Article number: 118379
SYMPTOMS
- System component (PDA detector) is not visible in the Run Samples window
- Connection of PDA displays as "ok? No" in node properties (Instruments tab)
- "System Idle - Instrument failure" appears on the bottom of the Run Samples window
ENVIRONMENT
- Empower 3 Enterprise
CAUSE
Interrupted or failed connection to PDA
FIX or WORKAROUND
- Power-down system components.
- Reboot the LAC/E.
- Restart the client and verify communication with the LAC/E.
- Turn on system components.
- Verify connection with all system components in node properties (Instruments tab) and in Run Samples.
ADDITIONAL INFORMATION