Empower projects and Review are very slow to open - WKB117541
Article number: 117541
SYMPTOMS
- Empower projects and Review are very slow to open
- Same issue when logged in to Empower directly on the database server
- Same issue for all users and all projects
- Very slow when switching between injections in Review
- The method set option "automatically apply method set" is not selected, and Review is set to start with the unnamed method set
ENVIRONMENT
- Empower 3 Feature Release 3
CAUSE
The message center was last purged almost four years ago and has nearly one million entries.
FIX or WORKAROUND
- Purge the message center to less than 300,000 entries.
- Monitor the message center to keep it at less than 300,000 entries.
ADDITIONAL INFORMATION
Empower 3 Service Release 3 introduces a new design for the message center that alleviates the need to manually purge or set up a scheduled purge.