Empower 3 MFC error extracting second wavelength chromatogram in Review - WKB117548

Article number: 117548

SYMPTOMS

  • After an Empower migration from Empower 2 to Empower 3, unable to extract a second wavelength chromatogram manually in the Review window
  • In all projects, this operation results in an MFC error, and the Review window closes unexpectedly
  • If working with only one wavelength, the application works as expected

ENVIRONMENT

  • Empower 3 FR5
  • Projects already created in Empower 2 FR5

CAUSE

Corruption in preferences

FIX or WORKAROUND

  1. In the project window, select file > Delete preferences of multiple views.
  2. Close the project window.
  3. Reopen the window and perform the operation that previously generated the error.

clipboard_e13ab072d77a4880fa936812e5c5a6080.png

ADDITIONAL INFORMATION

 