Agilent instrument panels are blank - WKB117176

Article number: 117176

SYMPTOMS

  • Instrument panels are blank in Empower
  • Status =No in node properties
  • Can ping the instrument LAN card

ENVIRONMENT

  • Empower 3
  • Agilent GC with or without Headspace
  • Agilent 6890N
  • Agilent 7890
  • Agilent G5664a

CAUSE

Wrong IP address in the PreConfiguration tool

FIX or WORKAROUND

  1. Use local host within the Agilent PreConfiguration tool.
  2. Reboot the LAC/E.