Error: "Data missing" in the injection review window for new and old projects and all users - WKB11655
Article number: 11655
SYMPTOMS
- "Data missing" error in the injection review window for new and old projects
- Same issue with all users
- In Empower > Configuration Manager > View > Manage RawData files > Test share fails
ENVIRONMENT
- Empower 3 Feature Release 3
- Citrix - client server
CAUSE
Windows security update removed the Projects, Network share path.
FIX or WORKAROUND
- Log into Empower > Configuration Manager > View > Click Manage RawData files and make note of the path.
- Open the drive where the raw is stored.
- Click Project and right-click project properties.
- Click the Sharing tab and add the path.