Lost communication with 2695 Separations Module and 2998 Photodiode Array Detector - WKB113347
Article number: 113347
SYMPTOMS
- Lost communication with 2695 Separations Module and 2998 Photodiode Array Detector
- IPv6 is disabled
- Empower folders are excluded from antivirus scan
- Message Center error: An error has occurred while attempting to request data, resulting in an instrument failure
- System Event Log: Network link is disconnected
ENVIRONMENT
- 2695 Separations Module
- 2998 Photodiode Array Detector
- Empower 3
CAUSE
Faulty instrument LAN card
FIX or WORKAROUND
Replace the instrument LAN card.