Is the Empower application supported with Oracle Transparent Application Failover (TAF)? - WKB113163
Article number: 113163
ENVIRONMENT
- Empower
- Oracle Transparent Application Failover (TAF)
ANSWER
No. Oracle Transparent Application Failover (TAF) should be set to disabled for the Empower application.
ADDITIONAL INFORMATION
This caused an issue in a customer environment where a transaction failed and had to roll back. CRI-2463 raised. Injections no longer associated with sample set. Error messages in message center:
- Oracle Error ORA-25405: transaction status unknown
- Oracle Error ORA-25402: transaction must roll back
- Oracle Error -2 trying to get next valid ID
Possibly related to Bug 21345368 - OCI Application Throws ORA-25405 From TAF And Transaction Guard Enabled Environment (Doc ID 21345368.8)