Spectra not available in Agilent G1315 A/B DAD detector - WKB111753
Article number: 111753
SYMPTOMS
- No spectra available in review window
ENVIRONMENT
- Agilent Instrument Control Framework v3.0 and above
- Agilent LC (G1315 A/B) DAD detector
- Empower Enterprise
CAUSE
The spectra range divided by the bandwidth is an uneven number.
Example:
Start wavelength 190 nm
End wavelength 400 nm
Slit 4 nm
Where, (400-190 = 210 divided by 4 = 52.5)
FIX or WORKAROUND
Based on the example calculation above, change either Start wavelength, End wavelength, or Slit to calculate as a whole number.
Example:
Where, (400-190 = 210 divided by 2 = 105)
ADDITIONAL INFORMATION
CRI-2436
