Skip to main content
Waters

How can I set the 2489 lamp to turn off during my Empower shutdown method? - WKB109493

Article number: 109493

OBJECTIVE or GOAL

Set the 2489 lamp to turn off during the Empower shutdown method.

ENVIRONMENT

  • Empower 3
  • 2489 Dual Wavelength Detector

PROCEDURE

  1. Open the instrument method being used for the shutdown method.
  2. Go to the section for the 2489 detector.
  3. Click on the Events tab.
  4. Put a check mark in the box for "Enable", found next to "Run Events".clipboard_e1e3b4e4e5620a4ca5c8e9d537df9fa77.png
  5. In row 1 of the events table, enter the time you want the lamp to shut off. Do not use 0 (0.1 is typically used).
  6. Set the "Event" field to "Lamp".
  7. Set the "Parameter" field to "Off".clipboard_e4c321c91a770bd33f3033b9be2188685.png
  8. Save the method.

ADDITIONAL INFORMATION

 

 

Not able to find a solution? Click here to request help.