Empower project backup process fails with the message "The export spawn process failed for project..." - WKB108972
Article number: 108972
SYMPTOMS
- Empower project backup fails with "The export spawn process failed for project..."
- EXP-00091: Exporting questionable statistics error messages are present in the Empower Backup Project Wizard window
ENVIRONMENT
- Empower 2
- Empower 3
CAUSE
An incorrect character set is defined in the registry for the string value NLS_LANG on the Empower node from which the backups are being executed. The default Empower data value for the string value NLS_LANG is AMERICAN_AMERICA.WE8ISO8859P1.
FIX or WORKAROUND
- Investigate why the value has changed.
- Edit the string value NLS_LANG and specify the expected Empower default value of AMERICAN_AMERICA.WE8ISO8859P1.
- Attempt to back up the project again.
ADDITIONAL INFORMATION
Caution! Using Registry Editor incorrectly can cause serious problems that might require you to reinstall Empower and/or the operating system. Waters cannot guarantee that problems resulting from the incorrect use of Registry Editor can be solved. Use Registry Editor at your own risk. Ensure that you back up the registry before you edit it.