In the Waters DHCP configuration window, the "Type" column is missing for ACQUITY SM-FTN and ACQUITY UPLC RI detector - WKB10860
Article number: 10860
SYMPTOMS
- The Type and Name columns are missing from the Waters DHCP Server Configuration window for some instruments
- IP and MAC addresses are present
- Removing IP addresses and rebooting the instruments does not resolve the issue
- All components are present for other Ethernet-controlled instruments
- Module is recognized with correct type in AutoLoader.exe
ENVIRONMENT
- Empower 3
- ACQUITY UPLC RI Detector
- ACQUITY Sample Manager - Flow Through Needle (SM-FTN)
- MassLynx
CAUSE
Possible corruption of instrument control software/drivers
FIX or WORKAROUND
- Remove all instrument control software/drivers.
- Reinstall only the necessary instrument control software for the components.
- Reboot the computer.
- Remove IP addresses from DHCP Service.
- Reboot instruments. Confirm that types appear in DHCP Service.