Skip to main content
Waters

Area and height counts between Thermo and Agilent GC are not the same in Empower - WKB108517

Article number: 108517

SYMPTOMS

  • Area and height counts between Thermo and Agilent GC are not the same in Empower

ENVIRONMENT

  • Agilent Instrument Control Framework
  • Empower 3 Enterprise
  • Agilent 7890 GC
  • Agilent 6890 GC
  • Agilent 6850 GC
  • Thermo GC

CAUSE

The Thermo driver uses a different conversion to report Area and Height in Empower.

FIX or WORKAROUND

Create a custom field for Thermo projects to match the value with respect to Agilent. 

Thermo_To_Ag_Area = Area*Scale to µV

Thermo_To_Ag_Height = Height*Scale to µV

ADDITIONAL INFORMATION

 

Not able to find a solution? Click here to request help.