Message center error: Chrom open for read failed, file \\HAGRID\Waters Projects$\7117\d28394.dat, err = 'Access is denied' - WKB107828
Article number: 107828
SYMPTOMS
- Batch processing on the LAC/E fails
- Message Center error: Chrom open for read failed, file \\HAGRID\Waters_Projects$\7117\d28394.dat, err = 'Access is denied.'
- Batch processing on the server is fine
- Bringing sample set into Review window and processing and saving are fine
- LAC/E and server are on the same domain
ENVIRONMENT
- Empower 3 Enterprise
- Feature Release 5
- Service Release 3
CAUSE
User access to fileshare is denied.
FIX or WORKAROUND
Change the login for the WatersFileService on the LAC/E to the Empower user account.
ADDITIONAL INFORMATION
id107828, LAC32, LAC32C07, LAC32C08, LAC32C09, LAC32C1_4, LAC32C10, LAC32C11, LAC32C12, LAC32C13, LAC32C15, LAC32C16, LAC32C5_6, LACE, LACE32C14