Error: Invalid Injection Volume, Requested 1µL does not match loop volume of 1000µL for Agilent 7890A GC and 7697B HS - WKB107578

Article number: 107578

SYMPTOMS

  • Error: Invalid Injection Volume, Requested 1µL does not match loop volume of 1000 µL for Agilent 7890A GC and 7697B HS
  • Agilent preconfiguration is set up properly
  • Able to set up instrument method in Run Samples window
  • Able to monitor in Run Samples window

ENVIRONMENT

  • Empower 3 Enterprise
  • Feature Release 4
  • Feature Release 5
  • ICF 3.1
  • ICF 3.4
  • Agilent 7697B Headspace
  • Agilent 7890 GC
  • Agilent 7697A Headspace
  • Agilent 8890 GC

CAUSE

The loop size installed on the HS is 1000 µL, but attempt is made to inject less than 1000 µL.

FIX or WORKAROUND

Set the injection volume in the sample set to 1000 µL. You can change the loop volume in preconfiguration to 0.001 mL if you want to set the injection volume to 1 µL.  However, the HS injection volume will be based on the installed loop volume. To inject less, you must physically install the preferred loop size and then pull in the HS configuration through the preconfiguration. Ultimately, the injection volume in the sample set must match the installed loop volume.

ADDITIONAL INFORMATION

 

id107578, 7890 GC, 7890 GCDET, 7890BGC, HS7697B