With Empower 3, is it possible to export the date and time on an exported text file? Is there a history? Is it possible to lock the raw data after export so that it cannot be tampered with? - WKB107283
Article number: 107283
ENVIRONMENT
- Empower 3
- Full Audit Trail
ANSWER
Is it possible to export the date and time on an exported text file?
No.
Is there a history (audit) for the export?
There is no audit remaining for export.
Is it possible to lock the raw data after export so that it cannot be tampered with?
The exported raw data is exported as a user-editable file; it cannot be exported with the locked state.
For tamper protection, please follow the SOP or consider performing the necessary calculations within Empower, without performing the export.