Article number: 10564

SYMPTOMS

  • After processing data, no results have amounts
  • No calibration curve is created

ENVIRONMENT

  • Empower 3
  • PDA

CAUSE

The data is being processed using a processing method only.

FIX or WORKAROUND

Reprocess data using a method set.

ADDITIONAL INFORMATION

In order to obtain amounts and a calibration curve, PDA data must be processed using a method set that contains a PDA processing method.