How to compare PDA spectra in review from several channels in Empower? - WKB10436
Article number: 10436
OBJECTIVE or GOAL
Compare PDA spectra in review from several channels in Empower. (Need to compare two spectra from two channels.)
ENVIRONMENT
- Empower 3
PROCEDURE
- Open in review the channels needed.
- Go to the menu Spectrum Review and click "Keep Spectra when Switching Channels".
- Right-click "Properties" on the spectrum plot and check "Overlay in Single Plot" in the Overlay tab.
ADDITIONAL INFORMATION