"Missing inject command" error message when trying to run samples in Empower with a CAD detector - WKB10233
Article number: 10233
SYMPTOMS
- "Missing inject command" error when trying to start a sequence in run samples
- Agilent 1200 and CAD detector both OK on the instruments tab of node properties
- Can control both the Agilent 1200 and the CAD detector through Empower
- Can set up and monitor a baseline with the CAD detector
ENVIRONMENT
- Empower 3
- Agilent 1200
- Corona Veo Charged Aerosol Detector (CAD)
CAUSE
An inject command is missing in the instrument method of the CAD detector.
FIX or WORKAROUND
- Add an injection command in the instrument method of the CAD detector.
- Run as normal.