In Empower 3, cannot process at the same time as running - WKB102031
Article number: 102031
SYMPTOMS
- Processing cannot be performed at the same time as running
- After running is finished, sample set is saved, but there is no result set
- Processing method is registered in the Method Set used for the run
- All sample set processes are "Normal"
ENVIRONMENT
- Empower 3
CAUSE
The Run Mode is set to "Run only".
FIX or WORKAROUND
Run the sample set with the Run Mode set to "Run and process" or "Run and report".
ADDITIONAL INFORMATION