Numeric custom fields return zero values - WKB101615

Article number: 101615

SYMPTOMS

  • Customer has two numeric custom fields:  A_Stdcurv_018 and B_Stdcurv_018
  • A_Stdcurv_018 = Peak type, Real; default value = 0.0000004653
  • B_Stdcurv_018 = Peak type, Real; default value = 0.008556
  • Another custom field uses these values: Konc_mg_ml_018
  • All return zero values when "Run and process" is used
  • Subsequent processing returns expected values

clipboard_e05ef5e887c7c8a5acd48d05ffc969f4e.png

clipboard_eab173851b2250d45e7adcf40d0ef6c29.png

ENVIRONMENT

  • Empower
  • Windows
  • Regional settings

CAUSE

Incorrect regional settings for the welcome screen (System Account). "Run and process" uses the SYSTEM account. 

FIX or WORKAROUND

  1. Ensure that the correct regional settings and decimal separator are set for the current user, the Welcome screen (System accounts), and New users.