Instrumentserver MFC application has stopped working - WKB1014
Article number: 1014
SYMPTOMS
- "Instrumentserver MFC application has stopped working" error message in Run Samples.
- Windows system and application logs show no obvious errors.
- No errors in message center.
- Virtual memory settings increased to recommended setting.
- Reboot of PC and systems do not resolve issue.
ENVIRONMENT
- Empower 3
CAUSE
Possibly a corrupt method or corrupt system
FIX or WORKAROUND
- Create new sample set method from scratch, not a "save as" method.
- If new method does not work create new system and check system configuration
ADDITIONAL INFORMATION
If you get an MFC application error, confirm the following:
- Ensure that your antivirus is not doing real time scans. Scheduled scans when the instrument is not running.
- Ensure that the Empower folder is excluded from antivirus scans.
- Ensure that you disable Windows automatic updates.
- Ensure that the following Windows services are disabled:
- Software Protection Service
- Windows Error Reporting
- Windows Security Center
To perform these steps and access the relevant services, you must have local administrative privileges on the computer on which you are working.
id1014, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP