Error "This <method name> is currently being edited by . . ." appears when locking a method in Empower - WKB100874
Article number: 100874
SYMPTOMS
- Error when locking or unlocking a method: This <Method Name> is currently being edited by . . .
ENVIRONMENT
- Empower 3 FR5 and higher
CAUSE
The Save Instrument Methods privilege for that method is missing.
In order to Lock or Unlock a method, the “Save” privilege for that particular method type is also required.
FIX or WORKAROUND
- With only the Lock Methods / Unlock Method privilege, a message similar to the following may appear:
- Add the appropriate method "Save" privilege for the user type in order to lock or unlock a method:
ADDITIONAL INFORMATION
id100874, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP