The e2695, 2489, and 2475 types do not show up in DHCP service, but IP addresses do - WKB10069
Article number: 10069
SYMPTOMS
- The e2695, 2489, and 2475 types do not show up in DHCP service, but IP addresses do show up
- 2489 and 2475 have instrument control software installed
- The instruments do not show up in loader.exe and do not show up on the Instruments tab in node properties
ENVIRONMENT
- Alliance Separations Module e2695
- 2489 UV/Vis Detector
- 2475 Fluorescence (FLR) Detector
- Empower
CAUSE
IP addresses are "stuck".
FIX or WORKAROUND
Reset the TCP/IP addresses. See the article How do I reset or clear the TCP/IP autoconfiguration address, 169.XX.XX.XX?