When viewing data or running samples, Empower reports a "fatal error" and then exits - WKB100373

Article number: 100373

SYMPTOMS

  • When viewing data or running samples, Empower reports a fatal error and then exits

clipboard_e1633b8c8b5cf14af7ac42a3ca31604af.png

ENVIRONMENT

Empower 3

CAUSE

System date and time was modified while Empower was running samples.

clipboard_ea46eb08dff8420d793a94a2a46353fcc.png

FIX or WORKAROUND

  1. Set the correct time. 
  2. Restart the computer.
  3. Empower should now work without errors.

ADDITIONAL INFORMATION