Cannot perform an injection with Beckman CE under Empower 3 control - WKB18200

Article number: 18200

SYMPTOMS

  • The Beckman CE appears to communicate with Empower
  • When the user starts the injection sequence, Empower never leaves "Waiting for Injection" status

ENVIRONMENT

  • Empower 3

CAUSE

The instrument method is not configured correctly.

FIX or WORKAROUND

In the instrument method, on the Time Program tab, the last Event line in the table should be "End".  Without the end line, the instrument will not inject.

ADDITIONAL INFORMATION