What happens if the gas pressure entering the Positive Pressure 96 Processor (186006961) does not reach 60 psi? - WKB80812
Article number: 80812
ENVIRONMENT
- Positive Pressure 96 Processor (PP96P)
- minimize flow variability
- manifold clamping mechanism
ANSWER
The minimum pressure ensures that 96 wells get adequate positive pressure to complete each step (so washes don't end up in the elution step solvent composition). It also ensures that all wells have equal pressure, to minimize flow variability to each sorbent bed.
ADDITIONAL INFORMATION
When minimum pressure is reached, it keeps the manifold clamping mechanism working effectively and consistently.
For additional information, see What is the minimum required gas pressure that can be used to operate part number 186006961 - Waters Positive Pressure - 96 Manifold?
id80812,