How much sample volume should I load on a 10-mg, 30-mg, 60-mg Oasis 96-well plate? - WKB54359
Article number: 54359
ENVIRONMENT
- Oasis solid-phase extraction
ANSWER
For the 10-mg sorbent, it is recommended to load 50-500 µL of sample.
For the 30-mg sorbent, it is recommended to load about 100 µL to 1 mL.
For the 60-mg sorbent, sample load is up to 2 mL.
ADDITIONAL INFORMATION
Loading capacity depends on total load based on mass, so both concentration and volume are important.