Skip to main content
Waters

What are the common causes of clogged wells in 96-well SPE plates? - WKB230669

Article number: 230669

ENVIRONMENT

  • Hydrophilic interaction chromatography (HILIC)
  • 96-well plates
  • GlycoWorks 
  • µElution
  • Oasis
  • Sep-Pak

ANSWER

There are many possible reasons, including:

  1. Sample viscosity is too high (dilute samples).
  2. Protein or other analyte has precipitated in the well, blocking flow.
    • Sirocco (filter) and Ostro (filter and SPE) plates are designed for in-well protein precipitation.
  3. Vacuum pressure too low (if using vacuum manifold).
    • Set vacuum pump to higher pressure.
    • Make sure the 96-well plate is sealing properly with the 96-well plate vacuum manifold.
    • Cover unused wells with lab tape to divert vacuum flow to wells that have the samples in them. 
  4. Positive pressure too low (if using positive pressure).
    • Make sure the pressure is set properly.
    • Make sure there is a good seal between the SPE plate and the positive pressure manifold.
    • Make sure none of the manifold positions above the wells is dirty or blocked.

Not able to find a solution? Click here to request help.